A203K Hipot/Flash/Dielectric Withstand Tester
Analogue Hipot/Flash/Dielectric Withstand Tester
The A203K/TL hipot/flash/dielectric withstand tester is a cost effective simple to use, rugged test system suited to manufacturing or laboratory environments. It can deliver a test voltage of 0 to 15 kV a.c. from an internal 500VA transformer source, and has displays that show both output voltage and current flow. Over current or flash/hipot/dielectric breakdown are indicated by a latching lamp and audible lamp, with separate fault reset button. This flash/hipot tester comes with the trademark Clare sprung shrouded safety HV probes.
Key Features
- Cost effective
- Simple to use
- Clear, easy to read, panel meters
- Rugged design
- Versatile
- Test probes supplied
- Adjustable voltage, 0-15 kV a.c. from a 500VA transformer source.
- Test voltage displayed on analogue meter scaled 0-15kV a.c. 5mA safety trip.
- Over current or flash-over dielectric breakdown indicated by maintained lamp and audible alarm.
- Fault indication reset by safety button.
- Complete with pair of sprung shrouded safety test probes.
Downloads
Please feel free to download any of the documents associated with this product.
Technical Specification
TEST OUTPUT VOLTAGE
Variable from 0-15 kV a.c.
TEST CURRENT TRIP RANGE
5mA fixed but can be bypassed for burn-out facility
DIMENSIONS
450L x 255H x 300D (mm)
WEIGHT
25 kg
SUPPLY VOLTAGE
120V or 230V
50 or 60 Hz